molbox1+ or molbox1+S v6.01 Embedded Software
TITLE | VER | USE WITH |
molbox1+/molbox1+S™ | 6.01 2016.12.16 |
molbox1+/molbox1+S |
TITLE | VER | USE WITH |
molbox1+/molbox1+S™ | 6.01 2016.12.16 |
molbox1+/molbox1+S |
This flash is intended ONLY for molbox1+ or molbox1+S with the Fluke Calibration logo:
on the front panel, serial number 2000 and above. Attempted use on any other molbox will result in an inoperable molbox unless the molbox1 hardware is upgraded to a molbox1+. For a molbox1 WITHOUT the Fluke Calibration logo, serial numbers 320 to 1500, please use the most recent |
||
Instructions: | ||
Download the file.
Run Winflash and select the file downloaded as the file to write to the product. Make sure the correct COM port of the product is connected to a RS232 port on the host PC. Use a null modem cable to COM2 port of PG7000 Piston Gauge. Use a standard RS232 cable to COM1 port of PPC, PPC, RPM and molbox. The updating of embedded software should only be performed after given specific instructions to do so from a Fluke Calibration Authorized Service Center ». If the update isn't performed properly, your product will be inoperable until it is. |
||
Updates Include: | ||
Rev 6.01 (2016 Dec 16) FIX (MBOX-367): Password entry screen corrupted in user level menu. Rev 6.00b (2011 Oct 11) Rev 6.00a (2010 Apr 15) |