Verifying the capacitance function of the 8588A Reference Multimeter

Document(s): 
application/pdf icon Verifying the capacitance function of the 8588A Reference Multimeter (1.21 MB(兆))

Overview: This application note provides guidance to those wishing to verify the performance of the capacitance function of their 8588A Reference Multimeter. It describes the methodology, examines the measurement process, and describes some of the pitfalls that can occur.

As capacitance is derived from resistance, do you need to routinely verify the capacitance function at all? To answer this, first look at how the 8588A Reference Multimeter measures capacitance.

The 8588A measures capacitance using a DC discharge method that is fundamentally the same as the method used by handheld multimeters, but of a higher precision. In the case of the 8588A, the unknown capacitor is discharged until it is close to zero and then charged with a known constant current; this results in a negative going linear ramp that returns to zero at the end of each measurement cycle. The voltages close to the start and end of the ramp are measured and the capacitance calculated from C = I.ẟt/ẟv where C is the capacitance in Farads, I is the current in amps and v is the change in voltage over the time interval t.

The resistance function provides the constant current and voltage measurement capability and uses the same calibration constants for voltage and current that were determined during resistance adjustment, except for the 1 nF range, which is adjusted using an external reference capacitor due to the difficulty in characterizing the internal bias currents.

The capacitance function is also affected by timing errors. Internally the 8588A has one master clock, so these timing errors can be evaluated by measuring the frequency counter. The clock specification is 1 µHz/Hz, which is many orders better than the specification of the best capacitance range, so any timing errors can be effectively ignored.

The Fluke Calibration 8588A Reference Multimeter

The 8588A measures 2 terminal capacitance using the DC discharge method

From this description you can see that the performance of the 10 nF range and above could be monitored by evaluation of the appropriate resistance range. For many laboratories this may be the preferred option due to lack of suitably traceable standards. It is possible that the capacitance measurement could be affected by internal physical changes that would not show as a change in resistance; however, any change would have a greater impact on the lower capacitance ranges, which can be evaluated by comparison with readily available reference capacitors. Any change seen on these lower ranges would very quickly become insignificant as you go up range.

To answer the question; do you need to separately verify the capacitance function? Yes.

For the 1 nF range because its calibration is independent of the resistance range it uses, but optional for the 10 nF range and above. This is because the performance of these ranges can be monitored by observing the change in the appropriate resistance range given in Table 1.

Capacitance range Resistance range used
1 nF 100 MΩ, Lo I
10 nF 10 MΩ, Lo I
100 nF 1 MΩ, Lo I
1 µF 100 kΩ, Lo I
10 µF 10 kΩ
100 µF 1 kΩ
1 mF to 100 mF 100 Ω
1 mF to 100 mF Lo I 1 kΩ

Table 1. Resistance to capacitance range mapping

Using reference capacitors to verify the 8588A

The value of a typical reference capacitor commonly found in calibration laboratories changes with frequency and its configuration as a 2-terminal or 3-terminal device. The 8588A measures 2 terminal capacitance using the DC discharge method described above. We therefore need to know what the effective test frequency of the 8588A is, and the 2-terminal value of the reference capacitor at this frequency.

There is no direct equivalence between the ramped waveform of the 8588A and the frequency of a sine wave. However, a good approximation can be made by equating the slew rate of the ramp generated by the 8588A with the maximum slew rate of a sine wave using the relationship: v/s =2.π.f.Vpk where v/s is the slew rate in volts per second, f is the frequency in Hz and Vpk is the peak voltage. Table 2 lists the equivalent test frequencies determined using this relationship. Note that the frequency not only changes with range but also where the value lies within the range.

8588A range Applied value % of range Equivalent test frequency
1 nF to 1 mF
 
20% 5.63 Hz
100% 1.13 Hz
200% 0.56 Hz
10 mF
 
20% 0.68 Hz
100% 0.14 Hz
200% 0.079 Hz
100 mF
 
20% 0.056 Hz
50% 0.023 Hz
100% 0.011 Hz
1 mF Lo I
 
20% 0.56 Hz
100% 0.11 Hz
200% 0.056 Hz
10 mF Lo I
 
20% 0.056 Hz
100% 0.011 Hz
200% 0.005 6 Hz
100 mF Lo I
 
20% 0.005 6 Hz
50% 0.002 3 Hz
100% 0.001 1 Hz

Table 2. 8588A effective test frequency

It is unlikely that traceability will be available at the frequencies listed in the table, so the reference capacitor will need to be measured at multiple frequencies going as low as possible and the value extrapolated.

Fortunately, the 1 nF range is much simpler to verify. Air dielectric capacitors are available at this value which have negligible frequency dependence; this allows them to be characterized at 1 kHz and used at these low frequencies. The main consideration with these capacitors is that they are generally 3-terminal devices. Therefore, they will need to be modified by fitting a shorting connector to their low terminal that connects it to the case and then measured as a 2-terminal device.

An alternative approach

A slightly different approach is used at Fluke due to the volume of instruments going through the laboratory. We use a characterized 5522A Multi-Product Calibrator that has been characterized using an 8588A transfer standard and the previously described 1 nF capacitor.

Technician Sets Up a Calibration Using the 8588A Reference Multimeter and the 5522A Multi-Product Calibrator

8588A and 5222A work together to create an alternate approach

The capacitance traceability for the transfer standard is derived from voltage, current, and time which has been validated against reference capacitors and a standalone measurement system which is also based on voltage, current, and time. This standalone measurement system has been used since the early 90’s by Fluke and is listed in the UKAS schedule of accreditation.

Keep learning

  • Using precision digital multimeters in place of analog null meters for metrological applications
  • Improved verification of a 55XX Series Multi-Product Calibrator with an 8588A Reference Multimeter
  • Three Surprising Features of the 8588A Reference Multimeter

See related products

  • Bench multimeters
  • 8588A Reference Multimeter
  • 5522A Multi-Product Calibrator

Get help

  • Customer and technical support
  • Speak with a calibration product expert about your equipment needs
  • 首页
  • 产品
    • 新产品
    • 电学计量校准
      • 标准器
      • 校准器
      • 高精度数字表
      • 电学校准软件
    • 射频校准
      • 射频参考标准
      • 射频校准配件
      • 射频校准软件
    • 数据采集和通用测试仪器
      • 台式万用表
      • 数据采集
      • 数据采集软件
    • 温度校准
      • 温度基标准
        • ITS-90 固定点容器
        • 标准铂电阻温度计 (SPRT)
        • 维修仪器
        • 液氮比较校准仪
        • 测温电桥
        • 标准电阻
      • 校准恒温槽
        • 袖珍型校准恒温槽
        • 标准型校准恒温槽
        • 特殊应用型恒温槽
        • 校准恒温槽配件
        • 校准恒温槽控制器
        • 校准恒温槽液
      • 工业温度校准仪
        • 现场计量炉
        • 计量炉
        • 手持干井校准仪
        • 现场干式校准
        • 微型恒温槽
        • 红外校准器
        • 热电偶炉
        • 双插块干井
        • 零点干井
      • 二等标准温度计/温度传感器
        • 铂电阻温度计 (PRT)
        • 热敏电阻
        • 热电偶
      • 测温仪
      • 热工多产品校准器
      • 温度校准系统及软件
    • 湿度校准
      • 湿度发生器
      • 湿度数据记录仪和监测仪
    • 压力计量校准
      • 基准级活塞式压力计
        • NMI 活塞式压力计
        • 绝压活塞式压力计
        • 高压力气动活塞压力计
        • 液压活塞式压力计
        • 活塞式压力计附件
      • 数字压力控制器/校准器
        • 低压控制器/校准仪
        • 气动压力控制器/校准仪
        • 高压气动控制器/校准仪
        • 液压控制器/校准器
      • 压力检测器
        • 低压检测仪
        • 数字压力计
        • 参考压力检测仪
      • 活塞式压力计
        • 气压活塞压力计
        • 油活塞压力计
        • 水活塞压力计
        • 高压液压自动测试仪
        • 活塞压力计附件附件
      • 手动压力校准及精密压力测量
        • 压力校准仪
        • 气压控制
        • 液压比较器/泵
      • Handheld Pressure Calibrators
      • 大气数据校准
      • 环境检测仪
      • 手动调压装置及增压泵
      • 定制压力校准系统
      • 压力校准软件
    • 流量校准
      • 气体流量标准
      • 气体流量配件
      • GFS动态质量流量基准
      • 流量校准软件
    • 过程校准仪器
      • 过程温度校准器
        • 手持式温度校准器
        • 干体式温度校准器和微型恒温槽
        • 精密数字温度计
        • 温度探头
        • 红外校准器
        • 带数据记录功能的温湿度仪
      • 压力校准器
        • 数字压力校准器
        • 手持式压力校准器
        • 活塞式压力计
        • 精密数字压力计
        • 手泵
      • 热工多产品校准器
      • 便携式过程校准器
      • 过程校准软件
    • 校准软件
      • MET/CAL® 软件
      • MET/CAL® 金牌支持
      • 资产管理软件
      • 温度校准系统及软件
      • 压力校准软件
      • 流量校准软件
      • 通用校准软件
    • 售后服务
    • 所有校准仪器
  • 售前咨询
    • 联系方式
    • 索取报价
    • 请求演示
    • 产品咨询
    • Certified Pre-Owned Equipment
    • General Services Administration (GSA)
    • Financing Program
    • National Stock Numbers (NSNs)
    • Payment Options and Tax Info
  • 新闻及动态
    • 新闻发布
    • 促销活动
    • Metrology Salary Survey
  • 培训与活动
    • Conferences and Exhibitions
    • 培训课程
    • User Group Meetings
  • 资料下载
    • Education Hub
    • 关于校准
    • 应用文章
    • 产品资料
    • 用户手册
    • 视频、音频及虚拟演示
    • Blog
    • 产品样本(简体中文)
    • 常见问题(简体中文)
    • 应用文章(简体中文)
    • 用户手册(简体中文)
  • 售后服务
    • Service Request (RMA)
    • Service Plans
    • 知识库
    • 认证
    • Calibration Certificates
    • Community Forum
    • My MET/SUPPORT
      • Activate
      • Procedures
      • Software
      • Technical Bulletins
      • Priority Support
      • Manuals
    • 产品手册
    • 安全数据表 (SDS)
    • 安全通告
    • 软件下载
    • 软件激活
  • 关于我们
    • 联系我们
    • 访问我们
    • 加入我们
    • 福禄克计量校准部简介
      • DH Instruments 压力和流量校准
      • Pressurements 压力校准
      • Ruska 压力校准
    • 福禄克其他部门
    • 商业准则
    • 常见问题
主页|文献|应用文章|电学计量校准|应用笔记|Verifying the capacitance function of the 8588A Reference Multimeter

在线客服
©1995-2022 Fluke Corporation

Secondary menu

  • Disclaimer
  • 隐私声明
  • 使用条款
  • Terms and Conditions of Sale