查看更多福禄克品牌
Fluke Calibration Blog
Calibration Equipment & Standards and Calibration Software
avr. 17, 2019
The best/proper method to clean your infrared calibrator target
The black body paint on the target of your infrared calibrator is very delicate. To remove the dust or debris from the paint, the only thing you can do is use compressed dry air. The dry air must have no oil and no water in the line or hose. Use a filter on an air hose to remove the oil and an air dryer to remove the water. Have the compressed air set to about 100 psi.
Some items like compressed air in a can will contain a...
avr. 16, 2019
If you are calibrating a medium to high accuracy RTD, you will probably use the characterization method of calibration. Characterization is the type of calibration in which the device under test (DUT) resistance is determined at several temperature points and the data are fitted to a mathematical expression.
To calibrate an RTD, perform these steps:
Place the reference probe and the DUTs in the temperature source. Make sure they are all placed as close together as possible, in a radial...
avr. 15, 2019
What knocks a digital instrument “out of calibration?”
Three common causes of calibration problems are:
Component shift. First, the major components of test instruments (for example, voltage references, input dividers, current shunts) can simply shift over time. This shifting is minor and usually harmless if you keep a good calibration schedule; this shifting is typically what calibration finds and corrects.
Drops. Suppose you drop a current clamp — hard. How do you know that clamp will...
avr. 12, 2019
Why Technicians Calibrate at the Bench -
Technicians calibrate at the bench to ensure calibrations are effective and don’t result in degradation of performance. They ensure that all components are in good working order prior to installation, and can evaluate them when component failure is suspected. The bench provides a stable ambient environment for calibration, an opportunity to use the most accurate equipment, and protection from factory conditions during the commissioning, testing,...
oct. 30, 2018
Thermocouples are common in industry because they are inexpensive and cover a wide temperature range. They should be tested during commissioning and again when removed from a process to verify that the tolerances were met. You’ll also want to test thermocouples at regular calibration intervals and if you suspect them of failing to meet performance specifications.
Because of their unique characteristics, it’s best to calibrate thermocouples in situ (in place) by comparison to a temperature...
juill. 24, 2018
Here’s how to do a simple comparison calibration of a probe-style RTD at one temperature point using temperature sensor calibration equipment, a dry-block calibrator in this case. Fully insert the probe RTD into one of the dry-block insert holes and connect the cable to an RTD temperature readout. Ramp the dry block to 100 ºC (212 ºF). When the dry-block display shows a stable 100 ºC, record the value of the RTD temperature readout. You now have one data point to compare...
juill. 24, 2018
Dry-block calibrators are stable temperature sources for calibrating sensors such as dial thermometers, resistance temperature detectors (RTDs) and thermocouples.These sensors are used for process manufacturing in many industries.They need to be checked or calibrated periodically. Watch the How to Calibrate a Digital Thermometer with a Dry Block video for an example.
Since dry blocks are portable, fast, and clean (do not use fluids), they are popular for on-site sensor...
mars 20, 2018
If you’re in the market for an isothermal bath (temperature calibration bath), you may have already noticed that many suppliers provide technical specifications that are confusing or incomplete. You might only be able to find a performance spec for a single temperature point, not for the bath’s whole temperature range. Or they don’t say which bath fluid was used in testing.
An isothermal calibration bath is a significant investment so you need to get enough information – and the right...
août 25, 2015
To calibrate an industrial temperature sensor like a PRT, RTD, thermistor, or thermocouple, you put the sensor into a stable industrial temperature readout like a dry-well, furnace, or calibration bath, and then compare the sensor’s output to the output of a reference standard probe connected to a thermometer readout. There are many different types of reference standard probes and thermometer readouts, and many ways to combine them, so knowing what to purchase can be a little confusing. We can...
Mai 19, 2014
Many Fluke Calibration pressure measuring instruments require periodic zeroing of the reference pressure transducer to correct for drift that occurs naturally over time. For the transducers that are intrinsically absolute, the zeroing process is often performed by comparison to local barometric pressure. This is a relatively simple task for laboratories that have access to a precision in-house barometer. With the exception of a few low pressure ranges, zero by comparison to an in...