I Just Got My Probe Recalibrated, Now What?

I just got my probe recalibrated, so now I just hook it up to the readout, load the coefficients and start measuring. I don’t have to worry about the probe’s calibration until it’s recalibrated.

Surprisingly, this is exactly how many people handle their probes’ calibrations. Following this practice saves a little time, but the cost can be severe. Recently, a company lost over a million dollars of product and six months of calibration work. The culprit was simply a PRT probe that shifted, probably during shipping, and wasn’t checked before being returned to service.

I hope this article will give you some ideas on how to make your temperature measurement program bulletproof by closing the loop on your probe calibrations.

What does “closing the loop” mean?

“Closing the loop” is completing the series of steps a probe should pass through as it is calibrated and used. In other words, there is more involved in managing a probe’s calibration than just sending the probe out for calibration every year.

The first step is a measurement taken to capture the status of the probe before it is boxed up and sent into shipping. This measurement closes the loop on the work the probe has done. When the probe is received back from calibration, another measurement is taken to verify that the probe didn’t shift during return shipment. Both the send-off measurement and return measurement are compared with the calibration report. This enables you to monitor the status of the probe, and it also provides a crosscheck between the calibration lab and your lab.

Measuring a probe before and after shipment may be the most important measurements in the overall calibration loop. Many probes and instruments are damaged beyond repair while in shipping. Make sure your work is closed out before shipping a probe.

How do I know if the probe is in tolerance?

Just because the probe was calibrated and given a calibration certificate doesn’t mean it was in tolerance. This might sound obvious, but I have had this conversation with enough customers to know that it is a common misconception. Someone who is surprised by this always follows up with the question “How do I know if the probe is in tolerance?”

Basically, the tolerance status of a probe is calculated by comparing the new calibration with the previous calibration. We answered this question in more detail in a previous Random News. Please refer to Random News Number 8.

How can “out of tolerance” be avoided?

The goal of all measurement programs is to make sure that everything always works and nothing is out of tolerance. Of course, the better the measurement program, the less often instruments are out of tolerance. Measuring a probe as it is sent out for calibration and when it is received from calibration is good, but the probe also needs to be monitored as it is being used.

One of the difficulties in measuring temperature is that probes are typically very fragile so their performance is very dependent on care and handling. Manufacturers may spec a probe’s long-term drift, but the only sure way to know the performance of a probe is by implementing interim checks and control charts (see the sample control chart in Figure 1). Whenever interim checks are mentioned, someone always says “Who has the time?” My answer is “Who has the time to recall six months or a year’s worth of work?” In the situation mentioned earlier, a 30-minute check every week would have saved the company over a million dollars and six months of work.

How often should the probe be checked?

There are no established rules that define how often an interim check should be done. One approach to establishing an interval is to decide the largest recall interval you are comfortable with and set the interim check interval accordingly. For example, if you calibrate 10 instruments a week with your temperature probe, then a weekly interim check may be sufficient. The largest risk would be the recall of 10 instruments. A daily interim check would limit the risk to recalibration of two instruments.

At each interim check, record the data on a spreadsheet or in a lab book. Graph the data. This is the beginning of a control chart. Add control lines to the chart to indicate how much the probe can drift before it exceeds the limits set in the calibration program (see Figure at right). The sample control chart shows the data of a probe measured about every seven days. On 4/28/06 the probe’s shift was verified to be about 0.006 °C, which exceeds the process limit. Only four days of cal work may have to be recalled due to the shift.

What temperature points should be measured?

This is an essential question to ask because, depending on the situation, many different temperatures can be used. Mainly, the temperature point(s) should be fairly simple to achieve and, importantly, the temperature point(s) must be repeatable. Let’s go over a few options.

The triple-point of water (TPW), 0.010 °C, is most often used for closing the loop in probe calibrations. At Fluke Calibration, the TPW is measured multiple times in every platinum resistance thermometer (PRT) calibration to monitor the probe as it is calibrated.

The TPW is very important in PRT calibrations because it can be used to troubleshoot several types of problems that affect PRT performance. Not only is the TPW included in every ITS-90 calibration, it is used in every ITS-90 measurement as well. The ITS-90 was designed to allow the user to update the probe’s RTPW (resistance at the triple-point of water). Updating the RTPW coefficient can also help reduce errors due to probe drift. For these reasons, the TPW is the first choice for a quick check to verify and close the loop on a PRT.

 

  • 首页
  • 产品
    • 新产品
    • 电学计量校准
      • 标准器
      • 校准器
      • 高精度数字表
      • 电学校准软件
    • 射频校准
      • 射频参考标准
      • 射频校准配件
      • 射频校准软件
    • 数据采集和通用测试仪器
      • 台式万用表
      • 数据采集
      • 功能和波形发生器
      • 数据采集软件
    • 温度校准
      • 温度基标准
        • ITS-90 固定点容器
        • 标准铂电阻温度计 (SPRT)
        • 维修仪器
        • 液氮比较校准仪
        • 测温电桥
        • 标准电阻
      • 校准恒温槽
        • 袖珍型校准恒温槽
        • 标准型校准恒温槽
        • 特殊应用型恒温槽
        • 校准恒温槽配件
        • 校准恒温槽控制器
        • 校准恒温槽液
        • 自定义校准槽
      • 工业温度校准仪
        • 现场计量炉
        • 计量炉
        • 手持干井校准仪
        • 现场干式校准
        • 微型恒温槽
        • 红外校准器
        • 热电偶炉
        • 双插块干井
        • 零点干井
      • 二等标准温度计/温度传感器
        • 铂电阻温度计 (PRT)
        • 热敏电阻
        • 热电偶
      • 测温仪
      • 热工多产品校准器
      • 温度校准系统及软件
    • 湿度校准
      • 湿度发生器
      • 湿度数据记录仪和监测仪
    • 压力计量校准
      • 基准级活塞式压力计
        • NMI 活塞式压力计
        • 绝压活塞式压力计
        • 高压力气动活塞压力计
        • 液压活塞式压力计
        • 活塞式压力计附件
      • 数字压力控制器/校准器
        • 低压控制器/校准仪
        • 气动压力控制器/校准仪
        • 高压气动控制器/校准仪
        • 液压控制器/校准器
      • 压力检测器
        • 低压检测仪
        • 数字压力计
        • 参考压力检测仪
      • 活塞式压力计
        • 气压活塞压力计
        • 油活塞压力计
        • 水活塞压力计
        • 高压液压自动测试仪
        • 活塞压力计附件附件
      • 手动压力校准及精密压力测量
        • 压力校准仪
        • 气压控制
        • 液压比较器/泵
      • Handheld Pressure Calibrators
      • 大气数据校准
      • 环境检测仪
      • 手动调压装置及增压泵
      • 定制压力校准系统
      • 压力校准软件
    • 流量校准
      • 气体流量标准
      • 气体流量配件
      • GFS动态质量流量基准
      • 流量校准软件
    • 过程校准仪器
      • 过程温度校准器
        • 手持式温度校准器
        • 干体式温度校准器和微型恒温槽
        • 精密数字温度计
        • 温度探头
        • 红外校准器
        • 带数据记录功能的温湿度仪
      • 压力校准器
        • 数字压力校准器
        • 手持式压力校准器
        • 活塞式压力计
        • 精密数字压力计
        • 手泵
      • 热工多产品校准器
      • 便携式过程校准器
      • 过程校准软件
    • 校准软件
      • MET/CAL® 软件
      • MET/CAL® 金牌支持
      • 资产管理软件
      • 温度校准系统及软件
      • 压力校准软件
      • 流量校准软件
      • 通用校准软件
    • 售后服务
    • 所有校准仪器
  • 售前咨询
    • 联系方式
    • 索取报价
    • 请求演示
    • 产品咨询
    • Certified Pre-Owned Equipment
    • General Services Administration (GSA)
    • Financing Program
    • National Stock Numbers (NSNs)
    • Payment Options and Tax Info
  • 新闻及动态
    • 新闻发布
    • 促销活动
    • Metrology Salary Survey
  • 培训与活动
    • Conferences and Exhibitions
    • 培训课程
    • User Group Meetings
  • 资料下载
    • Education Hub
    • 关于校准
    • 应用文章
    • 产品资料
    • 用户手册
    • 视频、音频及虚拟演示
    • Blog
    • 产品样本(简体中文)
    • 常见问题(简体中文)
    • 应用文章(简体中文)
    • 用户手册(简体中文)
  • 售后服务
    • Service Request (RMA)
    • Service Plans
    • 知识库
    • 认证
    • Calibration Certificates
    • Community Forum
    • My MET/SUPPORT
      • Activate
      • Procedures
      • Software
      • Technical Bulletins
      • Priority Support
      • Manuals
    • 产品手册
    • 软件激活
    • 安全数据表 (SDS)
    • 安全通告
    • 软件下载
    • 软件激活
  • 关于我们
    • 联系我们
    • 访问我们
    • 加入我们
    • 福禄克计量校准部简介
      • DH Instruments 压力和流量校准
      • Pressurements 压力校准
      • Ruska 压力校准
    • 福禄克其他部门
    • 商业准则
    • 常见问题
  • QQ客服在线咨询
主页|文献|应用文章|温度校准|文件/文章|I Just Got My Probe Recalibrated, Now What?

咨询福禄克计量校准产品的价格、销售渠道、技术等问题或申请产品的现场演示。

我们会尽快与您取得联系!

客服热线电话:400-8103435

©1995-2021 Fluke Corporation

Secondary menu

  • Disclaimer
  • 隐私声明
  • 网站管理员
  • 使用条款
  • Terms and Conditions of Sale